8.2 Functional Specifications
86
8.2
Functional Specifications
Display type
LCD
Display contents
Measured value, units (mA), recording (REC), endless recording (END-
LESS), statistical recording (STAT), recording interval (INTVL), date and
time (TIME), alarm (AL), battery status, recorded data count (DATA),
maximum value (MAX), minimum value (MIN), auto power saving (APS)
Operation button
Four ("SET", "REC/STOP", "+", "-")
Recording interval
1/2/5/10/15/20/30sec., 1/2/5/10/15/20/30/60 min.
Recording modes
• Instantaneous recording:
The instantaneous value is recorded at each recording interval
• Statistical recording:
Measurements are taken once per second, and instantaneous, maxi-
mum, minimum, and average values are saved at each recording inter-
val (cannot be selected when the recording interval is set to one
second).
Recording capacity
• Instantaneous recording: 60,000 values
• Statistical recording: 15,000 instantaneous, maximum, minimum, and
average values
Recording start method
• Logger button operation
• Instant or scheduled time (set by computer/Data Collector)
Recording stop method
• Logger button operation (endless recording)
• Logger button operation (one-time recording)
• Scheduled time (endless recording)
• Scheduled time (one-time recording)
Scheduled time is set by computer/Data Collector
Retained recording
sessions
Two sessions (each from recording start to stop)
Alarm
Indicates when measured values are outside of the range defined by
upper and lower thresholds set from a computer or the Data Collector
Scaling
Scales and displays measured values according to settings made from a
computer or the Data Collector
Power save setting
The measurement data display turns off about 30 seconds after the last
button operation (cancel power save for continuous display)
Real-time clock
Provided