5.1 Scaling
67
3
5
Chapter 5 Proces
si
ng Recorded Data
The following scaling calculation is applied to measured values.
Scaled Result = Raw data (measured value) × A + B × SI prefix (multiplier)
Scaled results are saved as a new item in the recording file.
1. Select the items, time span, and the following options.
* Set either one.
2. Confirm settings.
3. Click the
[Execute]
button.
(The scaled results are saved.)
Note: Click the
[Finish]
button to close the
[Scaling]
dialog box.
5.1
Scaling
Item and range settings
Select the item to be scaled, and the
time span.
Clicking this tab
changes the setting
options. Make set-
tings on either tab.
(The settings are ap-
plied to the other
tab.)
A/B (slope/offset) values
1
2
3
1
1
Setting Options
Description
Specify by
example *
Enter two known conversion points (up to ten digits each).
Specify by A/B *
Enter the scaling coefficients (A and B, up to ten digits each).
Scaled units
• Select the
[SI Prefix]
.
(
[p]
=1E-12,
[n]
=1E-9,
[
]
=1E-6,
[m]
=1E-3, blank =1E0,
[k]
=1E3,
[M]
=1E6,
[G]
=1E9,
[T]
=1E12)
• Enter a character string to identify the scaled units.
• (Up to five characters, except \, /, :, *, ?, ", <, >, and |.)
Setting
confirmation
Confirm that scaling is performed properly.Enter any numerical value as raw
data, and click the
[Calculate]
button to display the scaled result.