4.3 Starting and Stopping Recording
48
Depending on the selected recording interval, recording start is automatically delayed
until the next convenient clock time.
Automatic Recording Start at Convenient Times
Recording Interval
Recording Start Time
1 sec.
00 to 59 s (1-second interval)
2 sec.
00 to 58 s (2-seconds interval)
5 sec.
00 to 55 s (5-seconds interval)
10 sec.
000 to 50 s (10-seconds interval)
15 sec.
00 to 45 s (15-seconds interval)
20 sec.
00 to 40 s (20-seconds interval)
30 sec.
00 to 30 s (30-seconds interval)
1 min
00 min, 00 s to 59 min, 00 s (1-minute interval)
2 min
00 min, 00 s to 58 min, 00 s (2-minutes interval)
5 min
00 min, 00 s to 55 min, 00 s (5-minutes interval)
10 min
00 min, 00 s to 50 min, 00 s (10-minutes interval)
15 min
00 min, 00 s to 45 min, 00 s (15-minutes interval)
20 min
00 min, 00 s to 40 min, 00 s (20-minutes interval)
30 min
00 min, 00 s to 30 min, 00 s (30-minutes interval)
60 min
00 h, 00 min, 00 s to 23 h, 00 min, 00 s (1-hour interval)
At 12:01:00, you press the
REC/STOP
button.
The
[REC]
indicator in the display turns on,
but recording does not start yet.
At 12:10:00
Recording starts.
In the case of instantaneous value recording,
the number of data soon becomes 1.
In the case of statistical value recording, the
number of data becomes 1 at 12:20:00.
Example: When the button is pushed to start recording at 12:01:00,
and the recording interval is 10 minutes
12:01:00
Recording Start Operation
12:10:00
Actual Recording Start
If you try starting at 12:00 but are one minute late
Actual recording starts at 12:10, which is considered
the next convenient time.
www.
.com