Calibration
Calibrating the Voltage Measurement Unit
Use a generator that can output a DC voltage of 60 V.
Connect the instrument to the generator as shown below:
SENSE
SOURCE
SENSE
SOURCE
Model
BT3554
(-01)
DC generator
+
í
+
í
+
í
Do not apply AC current from the instrument to the generator, which may
cause the generator to malfunction.
8VHDORZRXWSXWLPSHGDQFH
ȍ
or less) generator.
If
[
íííí
]
is displayed, the disconnection detection function of the
instrument needs to be canceled.
Canceling the Disconnection Detection function
1
Turn off the instrument.
2
Press and hold the
A HOLD/MEMO
key while turning on the
power.
[on]
ÀDVKHV
3
Using the cursor keys, change the
[on]
to
[oFF]
.
4
Press the
ENTER
key.
This turns off the disconnection detection function and restarts
the instrument.
Restart the instrument after calibration. The disconnection
detection function will be turned on again. Do not cancel the
disconnection detection function during normal measurement.