(1) Noise Frequency Avoidance function
Functionality
Noise frequencies are avoided with the noise reduction
technology.
Default
OFF
How to set
Power-on option
Press and hold the
ȍ
key while turning the power on in
order to enter the setup mode.
Select from either on or off and press the
ENTER
key to
FRQ¿UPWKHVHWWLQJ
(Turning the power on again as usual, turns the function
off.)
(2) Zero adjustment function
Functionality
The measurement values obtained when the function is
performed (the correction values) are used as zero values.
Default
OFF
Correction
range
Up to 300 counts for each resistance and voltage range
How to correct
Press and hold the
0ADJ
key for 2 seconds to enter the
standby state.
The function obtains the correction values automatically
after
>íííí@
(for constant current fault detection or
disconnection detection) is canceled.
All correction values for all ranges is obtained at once.
The function exits (is disabled) if
[
íííí
]
is displayed for
10 seconds after it is started.
How to cancel
Press and hold the
0ADJ
key for 2 seconds when the zero
adjustment function is turned on, to cancel the function.