107
Error Messages
9.3 Error Messages
If the screen displays an error, follow the table below to troubleshoot
the error.
Message
Description
Solution
Err
Zero adjustment failed.
Connect the test lead correctly
before performing zero
adjustment. (p. 40)
Noise frequency avoidance
failed.
Connect the test lead correctly
during the measurement.
no.AdJ
Adjustment data error
The instrument has to be
repaired.
Contact your authorized Hioki
distributor or reseller.
Add.Err
A/D converter
communication error
Err01
Err02
Err03
Err04
Err05
Internal variable error
If
[
íííí
]
is displayed and
[OVER]
ÀDVKHVRQWKHVFUHHQDWWKH
VDPHWLPHWKHPD[LPXPGLVSOD\YDOXHVÀDVKHVWKLVGRHVQRW
indicate an error.
[
íííí
]
is displayed when input terminals have been opened.
Both
[OVER]
DQGWKHPD[LPXPGLVSOD\YDOXHÀDVKZKHQWKH
input values exceeds the set range. Set the range correctly.