Calibration
Appx. 6 Calibration
CAUTION
To prevent damage to the instrument, do not apply a voltage
between the positive (+) SOURCE and SENSE terminals or
EHWZHHQWKHQHJDWLYHí6285&(DQG6(16(WHUPLQDOV
Also, do not conduct measurements when the instrument is
turned off.
For the calibration environment, see the conditions of guaranteed accuracy (p.
LQWKH6SHFL¿FDWLRQVFKDSWHU
Calibrating the Resistance Measurement
Component
Use a standard resistor with little aging and with good temperature
performance.
Use a resistor with 4 terminals to prevent effects due to the resistor leads.
%HVXUHWRDVVLJQUHVLVWRUYD]$&IRUWKHLQVWUXPHQW8VLQJD
coil resistor will result in a higher inductance component. For this reason,
SXUH'&UHVLVWDQFHZLOOQRWHTXDOHIIHFWLYHUHVLVWDQFHWKHUHDOSDUWRI
impedance; displayed on the instrument).
Connect the instrument to the standard resistor as shown below:
SOURCE
SENSE
SENSE
SOURCE
Model
BT3554
(-01)
Standard
resistor
+
í
+
í