2.21 Electric Stretcher—Back Section Does Not Raise or Lower Electrically
Chapter 2: Troubleshooting Procedures
2.21 Electric Stretcher—Back Section Does Not Raise or Lower
Electrically
1.
Turn the crank handle to put the back section in a low to mid-height
position. This moves the unit off the limits of the potentiometer and will
make sure the up or down electrical operations work correctly.
2.
The stretcher has been connected to an external power supply and the
power cord is not damaged.
Yes
No
↓
→
Go to “Initial Actions” on page 2-1.
3.
On the nurse control, the
Head ON/OFF
control is in the
ON
position, and
the indicators are on (see figure 2-2 on page 2-9).
Yes
No
↓
→
4.
Push the
Head Up
or
Head Down
control on both siderails (see figure 2-1
on page 2-8). The head drive operates in both directions for both sides.
Yes
No
↓
→
5. Go to “Final Actions” on page 2-11
6. The plugs in the control board are all connected correctly.
SHOCK HAZARD:
Failure to unplug the stretcher from its power source could cause
personal injury or equipment damage.
CAUTION:
Failure to wear an antistatic strap when handling electronic components
could cause component damage.
Yes
No
↓
→
Unplug the stretcher from its power source. Put on an antistatic
strap, and then connect the plugs firmly into position. Go to step
2. If this solves the problem, go to “Final Actions” on page 2-11.
Otherwise, call Technical Support.
7. Go to “Final Actions” on page 2-11.
2
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 33
Service Manual (144386 REV 2)
Summary of Contents for P8000
Page 2: ......
Page 441: ...Schematic Wiring Diagram Control Board P N 49603 ...
Page 442: ...Schematic Wiring Diagram Electric Stretcher P N 60389 ...
Page 443: ......