2.26 OB/GYN Stretcher—Foot Supports Malfunction
Chapter 2: Troubleshooting Procedures
4.
Replace the defective foot support gas spring (refer to procedure 4.25). If
this solves the problem, go to “Final Actions” on page 2-11. Otherwise,
call Technical Support.
5.
The foot support (D) turns outward and inward when the release button is
engaged.
Yes
No
↓
→
Replace the defective foot support Mechlok®
device. If this
solves the problem, go to “Final Actions” on page 2-11.
Otherwise, call Technical Support.
6.
Go to “Final Actions” on page 2-11.
2
1. Mechlok® is a registered trademark of P. L. Porter Company.
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 39
Service Manual (144386 REV 2)