4.31 Electric Stretcher—Fuse
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.31 Electric Stretcher—Fuse
Tools required:
T25 Torx®
1
screwdriver
Removal
1. Raise the head section to the full up position.
SHOCK HAZARD:
Failure to unplug the stretcher from its power source could cause
personal injury or equipment damage.
2. Unplug the stretcher.
3.
Remove the two screws (A) that attach the cover (B) to the stretcher (see
figure 4-47 on page 4-93).
4. Remove the cover (B).
5. Remove the fuses (C).
Replacement
CAUTION:
Make sure the correct fuse is installed. Failure to do so could cause
equipment damage.
1. Install the new fuses (C).
2. Install the cover (B) on the stretcher.
3. Install the two screws (A) to attach the cover (B) to the stretcher.
4. Do the “Function Checks” on page 2-3.