4.10 Electric (P8020) Stretcher—Patient Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Replacement
SHOCK HAZARD:
Failure to unplug the stretcher from its power source could cause
personal injury or equipment damage.
1.
Clean any adhesive from the patient control unit (A).
2.
Remove the adhesive back from the new patient control switch.
NOTE:
The head symbol on the patient control switch should be at the head end of the
stretcher.
3.
Insert the wiring of the new patient control switch (B) through the patient
control unit (A), and install the switch into position.
4.
Connect the patient control switch (B) to the connector.
5.
Put the cover housing (C) into position, and install the two screws (D) to
attach it to siderail.
6.
Plug the stretcher into an applicable power source.
7.
Do the “Function Checks” on page 2-3.
Page 4 - 28
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)