7.25 Integrated Fiber Optic Exam Light (P7915AT)
Chapter 7: Accessories
Figure 7-29. Light Mounting Plate Assembly
19. Set the illuminator box (A) above the light mounting plate (B), so that the
power switch (C) is towards the foot end of the stretcher (see figure 7-26
on page 7-45).
20. Lower the illuminator box (A) onto the light mounting plate (B), and insert
the light mounting plate tabs (D) into the keyhole slots (E) located on the
bottom of the illuminator box (A).
21. Slide the illuminator box (A) toward the head end of the stretcher to secure
the illuminator box (A) onto the light mounting plate (B).
22. Remove the power cord (X) from its packaging container.
23. Plug the power cord (X) into the applicable connection on the illuminator
box (A).
CAUTION:
Make sure the power cord is not pulled too tightly. Tension on the power
cord connection could cause equipment damage.
24. Use a cable tie (F) to hold the power cord (X) on the light mounting plate
(B).
7
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 7 - 49
Service Manual (144386 REV 2)
Summary of Contents for P8000