Specific Function Checks—Surgical Stretcher
Chapter 2: Troubleshooting Procedures
Specific Function Checks—Surgical Stretcher
Back Section
1.
Lower the back section, and then use the left-side handle to raise the back
section to the high position. As you raise the back section, engage and
disengage the handle every 10° to 15°. The back section rises to the high
position, the gas spring releases and locks throughout its full range of
motion, and the handle engages and disengages completely. The handle
returns to its original position without binding.
NOTE:
The PACU extenders can either be removed from the stretcher, in the armboard
position, or in the extender position. If they are in the extender position
(around the articulating head section), then the handle on the PACU extenders
must be used to raise or lower the back section.
Yes
No
↓
→
2. Do step 1 for the handle on the right side of the stretcher.
Headrest
3.
Use the thumb push handle to raise and lower the articulating headrest
section, as well as turn in all directions (forward/backward, up/down). The
articulating headrest moves freely with no evidence of binding along both
axis. The gas spring releases and locks throughout its full range of motion,
and the push handle engages and disengages completely. The handle
returns to its original position without binding.
Yes
No
↓
→
4.
Go to “Final Actions” on page 2-11.
2
Specific Function Checks—Electric Stretcher
Patient Controls (located on the siderails)
1.
Raise the siderails, and lock them into position. Press and hold the
Head
Up
control (see figure 2-1 on page 2-8). The back section rises smoothly
without binding.
Yes
No
↓
→
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 7
Service Manual (144386 REV 2)
Summary of Contents for P8000
Page 2: ......
Page 441: ...Schematic Wiring Diagram Control Board P N 49603 ...
Page 442: ...Schematic Wiring Diagram Electric Stretcher P N 60389 ...
Page 443: ......