4.9 Electric (P8020) Stretcher—Left or Right Siderail Cable Assembly and Upright Tube
Chapter 4: Removal, Replacement, and Adjustment Procedures
Replacement
1.
Install the upper roller guide (C) and the plastic upper pivot (B) on the top
of the upright tube (D) (see figure 4-12 on page 4-24).
NOTE:
Make sure the wiring routing is correct.
2. Align the roller guide (I), the two wave washers (H)
,
and the upright tube
(D) with the bolt holes in the bottom rail.
CAUTION:
The lower pivot bolt has self-tapping threads. Make sure you use a new
lower pivot bolt for the installation procedure, and hand start it. Failure
to do so could cause damage to the bolt hole threads.
3.
Carefully install a
new
lower pivot bolt (J) to attach the tube to the bottom
rail.
4.
Tighten the lower pivot bolt (J) to 140 + 10 in-lb (15.8 + 1.1 N·m) of
torque.
5.
Align the upper roller guide (C) and the plastic upper pivot (B) with the
hole in the upper pivot.
6.
Install the two
new
ratchet rivets (A) to attach the upright tube (D) to the
upper pivot on the top rail.
7. Connect the wire connector at the patient control switch (R).
8.
Install the screw (M) and washer (P) with the cable clamp (Q) that attaches
the electric wire. Make sure the individual wires will not be seen once the
cover housing (O) is installed.
9.
Install the two screws (N) that attach the cover housing (O) to the patient
control switch (R).
10. Pull the siderail cable assembly (K) and connector through the strain relief
nut (Q) (see figure 4-4 on page 4-9).
11. Connect the left siderail wiring assembly connector (A) to P2 on the
control board (H) (see figure 4-17 on page 4-32).
12. Connect the right siderail wiring assembly connector (F) to P3 on the
control board (H).
4