4.17 Surgical (P8010) Stretcher—Articulating Headrest Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.17 Surgical (P8010) Stretcher—Articulating Headrest Gas Spring
Tools required:
½" open end wrench
Retaining ring removal/installation tool
Removal
1.
Loosen the jam nut (A) on the gas spring (B) (see figure 4-24 on page
4-48).
Figure 4-24. Surgical Stretcher Articulating Headrest Gas Spring
CAUTION:
Do not overextend the retaining rings. If the retaining rings are
overextended, use new ones for the installation. Failure to do so could
cause equipment damage.
2.
Remove the retaining ring (C) from the gas spring (B).
Replacement
1.
Raise the headrest, put the
new
gas spring (B) into position, and loosely
install the jam nut (A).
Page 4 - 48
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)