Assembly
Connect cooling water
Warning:
Risk of explosion
A closed valve in the recirculation system will build up high pressure in
the glassware and may result in glass bursting.
The pressure in the condenser shall not exceed 2 bar to ensure that a
higher pressure never occurs.
No valve or restriction may be built into the recirculation system.
Assembly parts:
2x tube nipples
(included in scope of delivery)
1x tube, inner Ø 7-8mm
(not included in scope of delivery)
Tube nipples
Input and output for cooling water is
located on the back of condenser.
Remove screw cap from input and
output of the cooling water.
Push screw cap over tube nipples so
that the black seal is placed inside the
cap.
Push intake and out ow hoses for
cooling water through screw cap over
one of each tube nipple.
Tighten screw cap with seal side up to
the openings of condenser.
Intake and out ow of cooling water are
connected.
Input and output cooling water