Assembly
Connecting peripheral devices
Connect vacuum
Warning / Attention:
Risk of implosion and poisoning/
Risque d’implosion et d’empoisonnement
Flasks which show even the slightest signs of damage may burst and
depending on the solvent poison the environment./
Les acons qui montrent des signes de dommage peuvent éclater et
en fonction du solvant peuvent empoissonner l’environnement.
Increase vacuum pressure and/or leakage pressure only if necessary.
Do not increase pressure/under-pressure beyond the limit required by
your solvents./
Augmenter la pression du vide et/ou la pression de fuite seulement si
cela est nécessaire. Ne pas augmenter la pression ou le vide au-delà
de la limite requise par vos solvants.
To connect the vacuum you need:
a vacuum source
a vacuum tube (Ø 7-8 mm inside)
Glassware set G1, G3 and G6
The opening of the vacuum connection is
located on the top of condenser
Unscrew either one of the screw caps
with nipple.
Push vacuum tubing onto nipple.
Tighten screw cap with nipple and
tubing.
Connect other end of vacuum tube
with vacuum source.
Screw caps vacuum
Vacuum connection G1, G3 and G6
Glassware set G5
The opening for vacuum connection is
located on the top of condenser.
Opening vacuum connection
Vacuum connection G5