EPS 45-80 High-Voltage Generator
3-24
Technical Manual P/N 9894.00.G6
3.4.3
RAD/Fluoro Frame Rate Selection
RAD and Pulsed Fluoro exposure rate can be selected either at the console or from the peripheral
equipment using up to three (3) programmable inputs (Figure 3.7 and Table 43).
•
The frame rate selection for RAD and Fluoro can either use the same lines or instead use a
distinct set of inputs. See Chapter 4, System Setup.
•
When remote frame rate selection is enabled, RAD and Fluoro are done at the operator
console's selection
when, and only when,
all frame rate selection lines are left open.
Figure 3.7
–
Typical Frame Rate Selection Inputs Interfacing.
Table 43-
Frame Rate Remote Selection Table
Bit 2
Bit 1
Bit 0
Frame Rate Operation
Open Open Open
Console
Selection
x
x
x
See Section 4.5.2 to program the frame rate selection