7. Troubleshooting & Servicing
Technical Manual P/N 9894.00.G6
7-29
Console Parameters
Use the Console parameters section to adjust any of the following setups.
a.
The Operator Console
Language
: Defaulted to English as main language and French as
second language. Note that French second language can be changed to most any language by the
client using the Roman, Greek or Cyrillic alphabet. Refer to Section
4.8.2.
b.
Contrast, Volume
and
Brightness
parameters: Adjustable for better operator console visibility
and listening. These parameters can also be set using the console service menu (see Section 4.8).
c.
mAs Copy Factor:
Used in console software version 9811.03 only. The "Copy Fluoro Factors"
function is available in
manual
Pulsed Fluoro. In that mode, depressing
F1
function key sets up
the RAD technique in two-point (mAs) mode with the kV equals to the actual Fluoro kV. The
RAD mAs parameter is a multiple of the actual Fluoro mAs; that factor can be modified by
depressing F1 key for a moment and by using the mAs Factor Setup menu or by using the “mAs
Copy factor” slider. The factor varies from 1.6 up to 10.0.
d.
“
Info Display
” selection: Used to default one of the six possible data to be displayed to the
operator in the Information Display area (Figure 7.16). This information can be changed at any
moment by the operator, using the left-hand third selection key (
I.D.A
. key) on the console
Figure 7.16
–
Graphic screen
on the Epsilon Operator Console
.
Different information can be
displayed in the
Information display
area using the
IDA
key.