1. Introduction to
Epsilon
HF
Generator
Technical Manual P/N 9894.00.G6
1-7
Auxiliary Power Supplies P/N 10023.000 (.002)
Provides a medical grade power supply rated 24 VAC, 150 VA (6.25A) and 24 VDC (2.7A), 65W.
Auxiliary Power Supplies P/N 10023.001
Provides a medical grade power supply rated 24 VDC (2.7A), 65W.
DAP Interface P/N 10028.000
The DAP Interface option allows interfacing a DAP measurement device
on the Epsilon Operator Console to display the radiation level a patient is
exposed to in radiological exams. Cumulative dose area product (
DAP
)
would also be printed when an optional DAP printer is connected to the
generator.
Cable Discharge Unit P/N 10027.000 (EPS CV configurations)
The
Cable Discharge Unit
provides the EPS 45-80 High-Voltage generator with an electronic crowbar
mechanism that permits multiple exposures at precise exposure times using rapid rise and fall times along a
broad spectrum of tube mA, on standard (no grid) X-ray tubes.
The quick turning on and off of the high-voltage output dramatically improves the quality of images acquired
from cineradiography, digital subtraction angiography, digital radiography and Pulsed Fluoro.
Optional
Cable Discharge Unit
(Shown here with fans for increased duty cycle)
Two typical kVp waveforms superimposed
for the benefit of comparison