EPS 45-80 High-Voltage Generator
4-40
Technical Manual P/N 9894.00.G6
4.8.2
Generator Clock
The generator real-time clock is set to GMT -05:00 Eastern Time (US & Canada). Use the History
Log module in Epsilon PC Tools to synchronize the generator clock with your service laptop.
1.
Connect the generator to the computer and open the PC Tools configuration file for the
generator.
2.
With Epsilon PC Tools in communication mode (Monitor/Transfer), select
Tools
/
Service/History Log
to invoke the History Log module (Figure 4.22).
3.
Set Date and Time
: Use this button to synchronize the generator clock with the PC.
Figure 4.22
–
PC Tools in
Service Mode
.
4.8.3
Language Module
The Operator Console
Language
is defaulted to English as main language and French as second
language. The selection of the console language can be made in service mode (Figure 4.23) or in
normal mode (using the User’s Monitor Mode, Section 7.5.2.8). Note that the second language can
be translated to any language by the client himself using the Roman, Greek or Cyrillic alphabet (See
Section 4.8.2.2).