California
Revision A
Instruments
INSTRUCTION MANUAL
4500L-3P-4728
Invertron
®
Page 1: ...California Revision A Instruments INSTRUCTION MANUAL 4500L 3P 4728 Invertron...
Page 2: ...ion This is accomplished by inhibiting the output voltage fault detection Before starting the calibration connect the test equipment as shown in Figure 1 The current shunt must be connected in series...
Page 3: ...1000 followed by pressing the PRG key and then the ENT key The following steps will calibrate the Current Limit function 1 Apply the load to the Phase A output terminal or the other phase to be calibr...
Page 4: ...1 RNG 04 2 Enter the key sequence 959 ENT 3 Depress the MNU key several times until the Configuration Menu Screen is displayed as illustrated below CFG 16 ALM 17 FLM 18 CLM 19 4 Enter the following ke...
Page 5: ...Figure 1 4...
Page 6: ...Figure 2 5...