3. 

Depress the MNU key several times until the first Menu screen is displayed as illustrated 
below: 

    

SNC 

01 

    

RNG 

04 

 

Enter the key sequence: 959 ENT 

4. 

Depress the MNU key several times until the Configuration Menu Screen is displayed as 
illustrated below: 

   CFG 

16 

 ALM 

17 

   FLM 

18 

 CLM 

19 

 

Enter the following key sequence to display the Configuration Screen: 

   16 

ENT 

5. 

While viewing the Configuration Screen, depress the B key and record the value. 

6. 

Enter the password recorded in step 2 followed by depressing the PRG key  Do not press the 
ENT key or the configuration value will be replaced by the password. 

7. 

Change the PHZ value to 1000 by pressing the C key and entering the value of 1000 
followed by pressing the PRG key and then the ENT key.  

The following steps will calibrate the Current Limit function. 

1. 

 Apply the load to the Phase A output terminal or the other phase to be calibrated. 

2. 

 Select the Current Limit (CRL) screen by pressing the following keys: 

   8 

ENT 

3. 

 Program the current limit to 10 amps with the following key sequence: 

   10 

PRG 

ENT 

 

Note: Only one output phase is displayed. All output phases will be programmed to the same 
value as phase A.  

4. 

Program the output voltage to 130 volts with the following key sequence: 

 

 

 

5 ENT 130 PRG ENT 

5. 

Check that the output current is 10.00 

±

 0.05 amps if not Slowly turn A9-R17, or the 

respective adjustment for the phase being calibrated, until the output current is 10.00 

±

 0.05 

amps. 

6. 

Repeat steps 1 through 5 for the other output phases. Refer to the following table for the 
current limit calibration adjustment. 

 