U10M, U10S
A1385-6.2 en/de/fr
HBM
5
Mechanical installation
5.1
Important precautions during installation
Handle the transducer with care.
Comply with the requirements for the force application parts as specified in
Chapter 5.3 and Chapter 5.5.
Welding currents must not be allowed to flow over the transducer. If there is
a risk that this might happen, you must use a suitable low‐ohm connection
to electrically bypass the transducer. HBM, for example, provides the highly
flexible EEK ground cable, which can be screwed on above and below the
transducer.
Make sure that the transducer cannot be overloaded.
WARNING
There is a danger of the transducer breaking if it is overloaded. This can
cause danger for the operating personnel of the system in which the
transducer is installed.
Implement appropriate safety measures to avoid overloads (see Specifica
tions in Chapter 9, Page 31) or to protect against resulting dangers.
5.2
General installation guidelines
The forces to be measured must act on the transducer as accurately as pos
sible in the direction of measurement. Torques and bending moments, eccent
ric loading and lateral forces may produce measurement errors and destroy
the transducer, if limit values are exceeded.
Summary of Contents for U10M
Page 2: ...English Page 3 42 Deutsch Seite 43 83 Fran ais Page 85 126...
Page 16: ......
Page 19: ......
Page 22: ......
Page 26: ......
Page 28: ......
Page 39: ......
Page 41: ......
Page 57: ......
Page 60: ......
Page 63: ......
Page 69: ......
Page 80: ......
Page 82: ......
Page 84: ...U10M U10S 84 A1385 6 2 en de fr HBM...
Page 99: ......
Page 102: ......
Page 105: ......
Page 112: ......
Page 123: ......
Page 125: ......
Page 127: ......