13
U10M, U10S
A1385-6.2 en/de/fr
HBM
4
Conditions on site
Protect the transducer from weather conditions such as rain, snow, ice and
salt water.
4.1
Ambient temperature
The temperature effects on the zero signal and on the sensitivity are com
pensated.
To obtain optimum measurement results, the nominal (rated) temperature
range must be complied with. The arrangement of the SG results in high
insensitivity to temperature gradients due to the construction. Nevertheless,
constant or very slowly changing temperatures are best. A radiation shield
and all‐round thermal insulation produce noticeable improvements, but must
not be allowed to set up a force shunt.
4.2
Moisture and corrosion protection
The force transducers are hermetically encapsulated and are therefore very
insensitive to moisture. The transducers reach the protection class IP67 per
DIN EN 60259 (test conditions: 0.5 hours under 1 m water column). Neverthe
less, the force transducers must be protected against permanent moisture
influence.
Optionally, the force transducers are also available fitted with a fixed cable. In
this version, all U10 achieve degree of protection IP68 with a nominal (rated)
force equal to or greater than 12.5 kN.
The transducer must be protected against chemicals that could attack the
steel (nominal (rated) forces from 12.5 kN) or aluminum (nominal (rated)
forces up to 5 kN).
With stainless steel force transducers, please note that acids and all materials
which release ions will also attack stainless steels and their welded seams.
Should there be any corrosion, this could cause the force transducer to fail. In
this case, appropriate protective measures must be implemented.
4.3
Deposits
Dust, dirt and other foreign matter must not be allowed to accumulate suffi
ciently to divert some of the measuring force onto the housing, thus invalidat
ing the measured value (force shunt). The relevant areas are marked in
Fig. 4.1.
Summary of Contents for U10M
Page 2: ...English Page 3 42 Deutsch Seite 43 83 Fran ais Page 85 126...
Page 16: ......
Page 19: ......
Page 22: ......
Page 26: ......
Page 28: ......
Page 39: ......
Page 41: ......
Page 57: ......
Page 60: ......
Page 63: ......
Page 69: ......
Page 80: ......
Page 82: ......
Page 84: ...U10M U10S 84 A1385 6 2 en de fr HBM...
Page 99: ......
Page 102: ......
Page 105: ......
Page 112: ......
Page 123: ......
Page 125: ......
Page 127: ......