99
This value is let out to the analog output and coincides
with the minimum value of the analog output type cho-
sen with setup item O .11 (analog output 1) and O .21
(analog output 2) .
• A new output value can be edited manually . It is possible
to change the value of the first blinking digit by pressing
the
or
key .
• To move to the next digit press the
key once; the first
digit will be fixed and the second one will start blinking .
• Pressing repeatedly the
key will result in cycling through
the digits on the primary LCD .
• Once a value is selected, the corresponding current is
immediately erogated by the selected output and no
confirmation is required .
• The minimum and maximum values let out are 3 .6,
22 mA for the 4-20 mA output and 0, 22 mA for the
0-20 mA output . This depends on the possibility of the
fault currents to be let out (see “Alarm - error configura-
tion” section for more details) .
• To verify the erogated current use a multimeter connected
to the corresponding output .
• To exit the test press the CFM key; the instrument will re-
main in the setup menu and move to the next setup item .
HOLD DIGITAL INPUT TEST
This test is made to verify if the instrument recognizes the
digital input signal at the hold input .
• To enter the test procedure select the setup item t .06 and
an “OFF” blinking message will appear on the LCD .
• Press the
(or
) key once and the message will switch
to a blinking “GO” .