31
Code
Valid Values
Default
Present
for ORP
P .03 Telephone number #2
(see notes 20, 27)
00000000000000000000 to
99999999999999999999
(“- - - - - - - - - - - - - - - - - - - -”
indicates no number)
- - - - - - - -
- - - - - - - -
- - - -
yes
P .04 Number of remaining mes-
sages (see notes 20, 28)
000 to 200 and 222
100
yes
P .05 Number of repeated SMSs
(see notes 20, 29)
0 to 5
2
yes
P .06 Delay between two subse-
quent SMSs (see notes 20, 30)
05 to 60 minutes
10
yes
P .07 SIM expiration day (see notes
20, 31)
01 to 31
01
yes
P .08 SIM expiration month (see
notes 20, 31)
01 to 12
01
yes
P .09 SIM expiration year (see notes
20, 31)
2000 to 2099
2010
yes
READING OFFSETS (“OFFS”)
F .00 pH or ORP actual value (see
notes 17, 25)
measured value -1 .00 pH or
-200 mV to measured value
+1 .00 pH or +200 mV
measured
value
yes
F .01 pH or ORP reading offset
adjustment (see note 25)
-1 .00 to +1 .00 pH or
-200 to +200 mV
0 .00 pH or
0 mV
yes
F .10 Temperature actual value (for
ATC only, see notes 18, 25)
measured value -10 .0 ºC to
measured value +10 .0 ºC
meas . value
(see note 17)
no
F .11 Temperature reading offset
adjustment (for ATC only, see notes
18, 25)
-10 .0 to +10 .0 ºC
0 .0 ºC
no
SOLUTION COMPENSATION (“SOLC”)
S .00 Solution compensation en-
able
“On”: compensation enabled
“OFF”: compensation disabled
“OFF”
no
S .10 pH value for point #1
-2 .00 to 16 .00 pH
7 .00 pH
no
S .11 Temperature value for point
#1 (SCT1)
-30 to 130 .0 ºC
|SCT1-SCT2| ≥1.0 ºC
20 ºC
no
S .20 pH value for point #2
-2 .00 to 16 .00 pH
7 .00 pH
no
S .21 Temperature value for point
#2 (SCT2)
-30 to 130 .0 ºC
|SCT1-SCT2| ≥1.0 ºC
30 ºC
no
IN-LINE CLEANING (“CLEA”)
SIMPLE CLEANING (“SCLE”)
L .00 Rinsing time
5 to 99 s
20 s
yes
L .01 Pause time
10 to 9999 minutes
1440 minutes yes