51
•
Rinsing time
: the device enters hold mode; all relays
configured for simple cleaning are energized . If the device
is in normal measurement mode, the “rinSinG” message
scrolls on the LCD; otherwise (i .e . the device is in setup
mode) the cleaning is performed, but no message ap-
pears .
•
Hold mode end delay
(set by item C .70): if the device
was controlling when the cleaning action started, then
the hold mode end delay must expire before restarting
control .
Advanced cleaning
actions are performed in the follow-
ing order:
•
Pre-rinsing time
: the device enters hold mode and relay
3 is energized . If the device is in normal measurement
mode, the “PrE-rinSinG” message scrolls on the LCD .
•
Detergent washing time
: relay 4 is energized (and relay
3 continues to be energized) . If the device is in normal
measurement mode, the “dEtErGEnt” message scrolls on
the LCD . It is possible to configure some cleaning without
detergent (setup item L .17) . In this case the message
displayed is “no dEtErGEnt” .
•
Rinsing time
: relay 4 is de-energized (and relay 3
continues to be energized) . If the device is in normal
measurement mode, the “rinSinG” message scrolls on
the LCD .
•
Hold mode end delay
: if the device was controlling
when the cleaning action started, then the hold mode
end delay must expire before restarting control .
The time each of these actions last for can be configured
through setup menu, items L .00 or L .10, L .11, L .12 .
When performing a cleaning action (either “simple “ or
“advanced”) with the device in normal measurement mode,
the secondary LCD displays a countdown for the seconds
remaining to the cleaning action end .