27
Code
Valid Values
Default
Present
for ORP
C .12 Hysteresis for setpoint 1 (H1)
(see note 1)
0 .00 to 18 .00 pH or
0 to 4000 mV
1 pH or
50 mV
yes
C .13 Deviation for setpoint 1 (D1)
(see note 1)
0 .50 to 18 .00 pH or
25 to 4000 mV
1 pH or
50 mV
yes
C .14 Reset time for setpoint 1
0 .1 to 999 .9 minutes
999 .9 min .
yes
C .15 Rate time for setpoint 1
0 .0 to 999 .9 minutes
0 .0 min .
yes
SETPOINT 2 (“SEt2”)
C .20 Setpoint 2 mode (M2)
(see note 1)
“OFF”: Disabled
“OOHI”: ON-OFF, high set point
“OOLO”: ON-OFF, low set point
“PIdH”: P .I .D ., high set point
“PIdL”: P .I .D ., low set point
“OOLO”
yes
C .21 Setpoint 2 (S2)
(see note 1)
-2 .00 to 16 .00 pH or
-2000 to 2000 mV
6 .00 pH or
-500 mV
yes
C .22 Hysteresis for setpoint 2 (H2)
(see note 1)
0 .00 to 18 .00 pH or
0 to 4000 mV
1 pH or
50 mV
yes
C .23 Deviation for setpoint 2 (D2)
(see note 1)
0 .50 to 18 .00 pH or
25 to 4000 mV
1 pH or
50 mV
yes
C .24 Reset time for setpoint 2
0 .1 to 999 .9 minutes
999 .9 mins
yes
C .25 Rate time for setpoint 2
0 .0 to 999 .9 minutes
0 .0 mins
yes
ALARM (“ALAr”)
C .30 Alarm relay delta value
for setpoint 1 (A1)
(see notes 1, 22)
0 .50 to 18 .00 pH
or
25 to 4000 mV
1 .00 pH
or
100 mV
yes
C .31 Alarm relay delta value
for setpoint 2 (A2)
(see notes 1, 22)
0 .50 to 18 .00 pH
or
25 to 4000 mV
1 .00 pH
or
100 mV
yes
C .32 Maximum relay ON time
(after that an alarm is generated)
1-60 minutes
60
yes
C .33 Alarm mask time
00:00 to 30:00 minutes
00:30
yes
PROGRAMMABLE CONTROL TIMING (“tIME”)
C .41 Hold time start
00:00 to 23:59 (see note 8)
00:00
yes
C .42 Hold time stop
00:00 to 23:59 (see note 8)
00:00
yes
C .51 Monday hold mode enable
“OFF”: Disabled
“On”: Enabled
“OFF”
yes
C .52 Tuesday hold mode
“OFF”: Disabled
“On”: Enabled
“OFF”
yes