32
Code
Valid Values
Default
Present
for ORP
ADVANCED CLEANING (“ACLE”)
L .10 Pre-rinsing time
0 to 99 s
20 s
yes
L .11 Detergent washing
0 to 99 s
10 s
yes
L .12 Rinsing time
5 to 99 s
20 s
yes
L .13 Pause time
Min . pause time (L .14) to
9999 minutes
1440 minutes yes
L .14 Minimum pause time
10 to Pause time (L .13)
10 minutes
yes
L .15 Cleaning trigger
“ ti”: Timer only
“ E”: Digital input or RS485
(external) only
“ti E”: Timer and digital input or
RS485 (external)
“tiEM”: Timer masked by digital
input (external)
“ ti”
yes
L .16 Repeated cleaning number
0 to 10
0
yes
L .17 Number of cleaning without
detergent
0 to 10
0
yes
TEMPERATURE LEVEL ALARM (“tEMP”)
U .00 Max . temperature level (see
note 32)
-30 to 130 °C
130 .0 °C
no
U .01 Min . temperature level
(see note 32)
-30 to 130 °C
-30 .0 °C
no
ERROR CONFIGURATION (“Erro”)
E .00 Alarm for setpoint 1 error
configuration
0 to 5
and 24 to 29 (see note 7)
3
yes
E .01 Alarm for setpoint 2 error
configuration
0 to 5
and 24 to 29 (see note 7)
5
yes
E .02 Max relay ON time error
configuration
0 to 5
and 24 to 29 (see note 7)
3
yes
E .03 Life check error configuration
(see note 11)
0 to 11
and 24 to 35 (see note 7)
9
yes
E .10 pH electrode broken or
leakage
0 to 11
and 24 to 35 (see note 7)
9
no
E .11 Reference electrode broken
or dirty or not immersed
0 to 47 (see note 7)
21
yes
E .12 Old pH probe error configu-
ration
0 to 11
and 24 to 35 (see note 7)
0
no
E .13 Dead pH probe error con-
figuration
0 to 11
and 24 to 35 (see note 7)
2
no