29
Code
Valid Values
Default
Present
for ORP
0 .12 Measurement value for ana-
log output 1 minimum
(0_VARMIN1)
-2 .00 to 16 .00 pH
or -2000 to 2000 mV
(0_VARMIN1 ≤ 0_VARMAX1-1
pH or 50 mV,
0_VARMIN1 ≤ 0_HOLD1≤
0_VARMAX1)
0 .00 pH or
-2000 mV (6)
yes
0 .13 Measurement value for ana-
log output 1 maximum
(0_VARMAX1)
-2 .00 to 16 .00 pH
or -2000 to 2000 mV
(0_VARMIN1 ≤ 0_VARMAX1-1
pH or 50 mV,
0_VARMIN1 ≤ 0_HOLD1≤
0_VARMAX1)
14 .00 pH or
2000 mV (6)
yes
0 .14 Analog output 1 value upon
hold mode (see note 16)
“USEr”: User selected value
“HOLd”: Previous value is frozen
“HOLd”
yes
0 .15 Analog output 1 value
upon hold mode if 0 .14=”USEr”
(0_HOLD1)
-2 .00 to 16 .00 pH
or -2000 to 2000 mV
(it must be within the 0_VARMIN1
to 0_VARMAX1-1 interval)
7 .00 pH or
0 mV (6)
yes
ANALOG OUTPUT #2 (“Out2”)
0 .20 Analog output 2 mode
(see notes 21, 26)
“rECO”: Recorder (temperature)
“SEt”: Control, setpoint 2 (pH) or
“OFF”: Disabled
“SEt”: Control, setpoint 2 (ORP)
“rECO” (pH)
or
“OFF” (ORP)
yes
0 .21 Analog output 2 type
“0-20”: 0-20 mA
“4-20”: 4-20 mA
“4-20”
yes
0 .22 Measurement value for ana-
log output 2 minimum
(0_VARMIN2)
-30 to 130 .0 ºC
(0_VARMIN2 ≤ 0_VARMAX2-10
ºC, 0_VARMIN2 ≤ 0_HOLD2≤
0_VARMAX2)
0 ºC
no
0 .23 Measurement value for ana-
log output 2 maximum
(0_VARMAX2)
-30 to 130 .0 ºC
(0_VARMIN2 ≤ 0_VARMAX2-10
ºC, 0_VARMIN2 ≤ 0_HOLD2≤
0_VARMAX2)
+100 .0 ºC
no
0 .24 Analog output 2 value upon
hold mode (see note 16)
“USEr”: User selected value
“HOLd”: Previous value is frozen
“HOLd”
no
0 .25 Analog output 2 value
upon hold mode if 0 .24=”USEr”
(0_HOLD2)
-30 to 130 .0 ºC
(it must be within the 0_VARMIN2
to 0_VARMAX2-1 interval)
25 ºC
no
BAUD RATE (“bAud”)
0 .30 Baud rate (see note 20)
1200, 2400, 4800, 9600,
19200
19200
yes