52
If the “Repeated cleaning number” value (setup item L .16)
is different from 0, then the advanced cleaning is repeated
a number of times equal to this value after the first cycle
(e .g . one more cycle if L .16=1) .
Note
If a cleaning session is being performed, it is possible to
stop it by pressing and holding the
and
keys together
(
key first) or through RS485 by sending the appropriate
command . When the advanced cleaning is aborted, in
any case a complete rinsing phase is performed before the
actual termination of the cleaning action . If the abortion
is issued during the rinsing phase, the phase continues
normally till its natural end .
Note
Calibration of whatever input or output can not be initiated
while the Simple or Advanced Cleaning is in progress .
Conversely, cleaning can not be triggered while whatever
calibration is being performed .