33
Code
Valid Values
Default
Present
for ORP
E .14 Calibration time-out error
configuration
0 to 11
and 24 to 35 (see note 7)
0
no
E .20 Temperature broken probe
error configuration
0 to 11
and 24 to 35 (see note 7)
3
no
E .21 Temperature level error
configuration
0 to 5
and 24 to 29 (see note 7)
3
no
E .40 Digital Transmitter error
configuration
0 to 5
and 24 to 35 (see note 7)
9
yes
E .50 Cellular error configuration
0 to 11 (see note 7)
3
yes
E .90 Power reset error configura-
tion
0 to 11
and 24 to 35 (see note 7)
2
yes
E .91 EEPROM corruption error
configuration
6 to 11
and 30 to 35 (see note 7)
9
yes
E .92 Watchdog error configuration 0 to 11
and 24 to 35 (see note 7)
2
yes
E .99 Level or pulse alarm relay
signal (see note 13)
“LE”: Level
“PULS”: Pulse
“LE”
yes
TEST (“tESt”)
t .00 Display test
“OFF”: To skip without testing
“GO”: To start the display test
“OFF”
yes
t .01 Keyboard test
“OFF”: To skip without testing
“GO”: To start the keyboard test
“OFF”
yes
t .02 EEPROM test
“OFF”: To skip without testing
“GO”: To start the EEPROM test
“OFF”
yes
t .03 Relays and LEDs test
“OFF”: To skip without testing
“GO”: To start the test
“OFF”
yes
t .04 Analog output 1 test
“OFF”: To skip without testing
“GO”: To start the test
“OFF”
yes
t .05 Analog output 2 test
“OFF”: To skip without testing
“GO”: To start the test
“OFF”
yes
t .06 Hold digital input test
“OFF”: To skip without testing
“GO”: To start the test
“OFF”
yes
t .07 Advanced cleaning digital
input test
“OFF”: To skip without testing
“GO”: To start the test
“OFF”
yes
Warning: cleaning actions could be triggered by the input going high during the test (see note 23) .
t .08 Watchdog test
“OFF”: To skip without testing
“GO”: To start the watchdog test
“OFF”
yes