87
Pressing the
key, the meter will cycle through the follow-
ing steps in reverse order, i .e . beginning from last buffer .
Note
At any time pressing LCD or CAL DATA key the meter will
return to the regular operating display .
• Press the
or
key to view the
time of last calibration . The second-
ary display will show “HOU” to
indicate “hour and minute” while
the primary LCD will show the time
as HH:mm format .
• Press the
or
key again to
view the offset in mV at the time
of last calibration . The secondary
display will show “OFF” to indicate
“offset” .
• Press the
or
key again to view
the slope in mV/pH at the time of
last calibration . The secondary
display will show “SLO” to indicate
“slope” .
• Press the
or
key again to view
the first memorized buffer at the
time of last calibration . The sec-
ondary display will show “BUF1”
to indicate “first buffer” .
• Press the
or
key again to view
the second memorized buffer (if
present) at the time of last calibra-
tion . The secondary display will
show “BUF2” to indicate “second
buffer” .
• Press the
or
key again to return
to the first CAL DATA display (date)
at the time of last calibration .
Note
When the “Measurement input selection” item is set to
“Digital Transmitter” the displayed last calibration data
refer to the Digital Transmitter and is stored in that device .
This does not mean that the calibration data of the process
controller are lost, but they come back when changing back
the “Measurement input selection” later .