CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
Chapter 6, Troubleshooting
6-25
SOFTWARE
WATCHDOG
FAULT - CURRENT
MODULE ID=X
EXPECTED
MODULE ID=X
CURRENT
STATE=X
SOFTWARE
WATCHDOG
FAULT - INVALID
STATE=X
SPM / DPM FILTER
IS NOT
INSTALLED -
PLEASE INSTALL
AND UNCLAMP IT
SPM FILTER IS
NOT INSTALLED -
PLEASE INSTALL
AND UNCLAMP IT
THE PROTOCOL
PARAMETERS
HAVE RESET TO
THEIR ORIGINAL
VALUES
CPU did not execute
required steps
resulting in watchdog
time-out.
CPU did not execute
required steps
resulting in watchdog
time-out.
DPM and SPM
interlock signals
indicate that DPM
and SPM filters are
not installed anytime
after system
initialization.
SPM interlock
signals indicate that
SPM filter is not
installed during Draw
or Return.
Protocol parameter
CRC does not match
value stored in
NVRAM.
1. Replace the processor card. Using Utilities, calibrate all
the components under A/D, centrifuge, safety system,
and configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace the processor card. Recalibrate A/D, centrifuge
and safety system, and reconfigure the language in
Utilities. Reset protocol parameters (MODIFY key), if
parameters were different from default values.
1. Verify that white sleeves are fully seated.
2. Check, repair, or replace all electrical connections
associated with the DPM and SPM.
3. Replace DPM and SPM. Use Utilities to calibrate DPM
and SPM in A/D screen, and safety system screen for
DPM only.
4. Replace the processor card. Using Utilities, calibrate all
the components under A/D, centrifuge, safety system,
and configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Verify that white sleeves are fully seated.
2. Check, repair, or replace all electrical connections
associated with the SPM.
3. Replace SPM. Use Utilities to calibrate SPM in the A/D
screen.
4. Replace the processor card. Using Utilities, calibrate all
the components under A/D, centrifuge, safety system,
and configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Normally displayed the first power-up after a program
chip has been changed. If message is reoccurring,
replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.