6-28
PCS2 Service Manual
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
Please Do Not Touch
The Weigher
Unexpected Weight
Change
Centrifugation
Interrupted With A
Full Bowl
Draw Is Not Allowed
Press RETURN to
continue
Return Cycle
Aborted
*The Bowl Is Not
Empty
Press RETURN to
continue
XXXXX Warning
XXXXX
New Cycle Will Start
Press NO to Continue
Return
Press YES to start
New Cycle
The machine detects
a rapid change in
weight.
During the beginning
of the first Draw, the
machine detects the
weight is not within
its tare range.
PCS2 senses a full
bowl and will not
allow Draw to
continue to prevent
red cell spillage.
PCS2 senses a bowl
partially full of
packed red cells and
does not suggest
proceeding with
Draw to prevent red
cell spillage.
Confirmation screen
to Acknowledge that
a new Cycle will
begin. Machine has
recommended that
the bowl be emptied.
1. Ensure weigher arm is securely fastened to the load cell
upper standoff and the lower standoff rotates as the arm
is rotated.
2. Ensure weigher arm is aligned perpendicular to the PCS2
when fully extended.
3. Verify the weigher calibration in Diagnostics. Re-
calibrate if necessary. See Weigher under A/D in Chapter
4, Calibration.
1. Ensure collection container is properly placed on the
weigher arm.
2. Ensure collection container is not touching the side of the
machine.
3. Press YES to resume the procedure. The PCS2 will tare
the collection container on the weigher arm again.
Press RETURN to continue the procedure.
Note:
If the Draw sequence was interrupted BEFORE the
plasma collection started, the Draw sequence may be
resumed manually as soon as it is safe to do so. In this case
the PCS2 will not post this warning.
Press RETURN to continue the procedure. (The bowl has to
be emptied of packed cells to ensure a properly controlled
collection.)
Press RETURN to continue the procedure. (The bowl has to
be emptied of packed cells to ensure a properly controlled
collection.)