Chapter 5, Installation and Configuration
5-7
6. Remove the weights and verify that the displayed weight returns to
ZERO ±1.
7. If calibration is required, refer to Chapter 4.
8. Exit the weigher test by pressing the MODIFY PROGRAM key.
DPM Sensor Test
1. Be sure no filters are connected to the DPM and SPM. Select the
Pressure test using the MODIFY PROGRAM and DRAW keys.
Note:
List Number 06002-110-NA machines do not have the SPM.
2. Verify that the DPM offset is within the range of 1984 to 2112. If out
of range, DPM may need to be replaced.
3. Grasp the DPM white luer plunger and twist counterclockwise while
sliding in the plunger. Verify both interlock switch status indicators
change to IN. Release the plunger and verify that both indicators
return to OUT.
4. Grasp the DPM white luer plunger and twist clockwise while sliding
in the plunger. Verify both interlock switch status indicators change to
IN. Release the plunger and verify that both indicators return to OUT.
5. Install a DPM filter with about 4" of tubing onto the DPM. Connect a
calibrated sphygmomanometer or equivalent device to the filter
tubing. Pressurize the DPM to 300 mmHg and verify that the
displayed pressure reading is within the range of 292 to 308 mmHg.
6. If calibration is required, refer to Chapter 4.
Cuff Test
1. Install a cuff adaptor tubing with a sphygmomanometer (or
equivalent) attached into the rear panel cuff connector.
2. Turn the cuff compressor on by pressing the CUFF key.
3. Verify the cuff inflates to 50 ± 5 mmHg.
4. Use the YES key to increase the cuff pressure to 100.
5. Verify the cuff inflates to100 ± 10 mmHg.
6. If calibration is required, refer the Chapter 4, Calibration.
Exit the Pressure test by pressing the MODIFY PROGRAM key.
Display Test
1. Select the Display test using the MODIFY PROGRAM and DRAW keys.
2. Verify that all pixels are lit on display, then press DRAW.
3. Verify that a character set is displayed and no characters are deformed
or missing.
4. Exit the Display test by pressing the MODIFY PROGRAM key.