CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
Chapter 6, Troubleshooting
6-7
13
15
16
DONOR VALVE
CURRENT FAULT
PLASMA VALVE
CURRENT FAULT
SALINE VALVE
CURRENT FAULT
WATCHDOG
ERROR
WATCHDOG
ERROR
Donor valve current
sense circuit indicates
that current did not
transition high or
low.
Plasma valve current
sense circuit indicates
that current did not
transition high or
low.
Saline valve current
sense circuit indicates
that current did not
transition high or
low.
Safety system relay
did not open during
system start-up
watchdog test.
Safety system relay
did not close during
system start-up
watchdog test.
1. Use the valve test in Diagnostics to manually transition
valves and monitor the valve and switch response.
2. Swap valve connections at the top deck distribution card
to isolate problem to valve or circuit.
3. Replace driver card.
4. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace driver card.
2. Replace safety card.
3. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace driver card.
2. Replace safety card.
3. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.