Chapter 5, Installation and Configuration
5-15
PCS2 Machine Transfer Verification Form
Customer Name:
City/State:
PCS-2 Serial #
VALVES
Functional
Yes
No
WEIGHER
Zero Reading
0 gm ± 1
1000 gm ± 5
PRESSURE
DPM
Offset (1984 - 2112)
300 mmHg ± 8
CUFF
50 mmHg ± 5
100 mmHg ± 10
DISPLAY
Functional
Yes
No
KEYBOARD
Functional
Yes
No
CENTRIFUGE
SPEED
Set Speed
Actual Speed
7000 RPM ± 100
Cover Lock
Functional
Yes
No
Bowl Vibration
Functional
Yes
No
Spill Detect
Functional
Yes
No
Donor Lights
Functional
Yes
No
TRANSDUCERS
OPTICS
Aim (Yes if Adjusted)
Yes
No
Open (1-82 or 0 V ± 0.10)
Grey Card
(2295- 3278 or 2.8 - 4.0 V)
LINE SENSOR
Open (2866 - 3276)
Black (0-103)
AIR DETECTOR
ACAD
Functional
Yes
No
BLAD
Functional
Yes
No
DLAD-1
Functional
Yes
No
DLAD-2
Functional
Yes
No
FANS
Functional
Yes
No
Biohazardous Waste Bag Attached
Yes
No
Program Revision
Cuff
Weigher
Circle if adjusted
DPM
Centrifuge
Optics Voltage
Line Sensor
Performed by:
Print Name:
Date: