2-24
PCS2 Service Manual
7. Remove the (2) Phillips head screws securing the power supply
assembly to the cabinet.
8. Pull the power supply assembly out the rear cabinet cavity by lifting
the rear edge and pulling the front edge away from the plastic edge
clips on the cabinet base.
9. Remove the two (2) Phillips head screws securing the left hand
bracket to the power supply.
10. Remove the four (4) Phillips head screws securing the right hand
bracket to the power supply.
11. Slide the power supply away from the power supply mounting frame.
12. Recalibrate the Condor power supply referring to the procedure in
Chapter 4, Calibration.
Centrifuge Controller Card
1. Remove front and rear panels. (See page 2-2.)
2. Disconnect P1, P2, and P3 from the centrifuge controller card.
3. Remove the (4) Phillips head screws securing the centrifuge controller
card to the power supply assembly cage.
4. Recalibrate the centrifuge referring to the procedure in Chapter 4,
Calibration.
1. Remove the rear panel. (See page 2-2.)
2. Remove the (2) Phillips head screws securing the AC or saline pole
holder assembly to the cabinet.
1. Tilt the cabinet and remove the (4) Phillips head screws securing the
spill bag plate to the base of the cabinet.
2. Carefully cut the tubing from the drain tube.
3. Replace with new drain tube assembly.
1. Remove the rear panel. (See page 2-2.)
2. Disconnect P405 from the centrifuge distribution card.
3. Remove the (2) fiber-optic cable ends from the photoelectric assembly
by unscrewing the (2) plastic thumb screw fasteners.
4. Remove the (2) Phillips head screws securing the photoelectric
assembly to the power supply assembly cage.
Note:
When reinstalling fiber-optic cable ends into the photoelectric
assembly, tighten plastic fasteners securely. Refer to Chapter 4,
Calibration, to adjust the bowl optics signal.
Photoelectric
Assembly
AC and Saline
Pole Holder
Assembly
Drain Tube
Assembly