Operating
46
Example:
Set value programmed at the circulator
T
set
= 70.7
°
C
Actual temperature in bath / system
T
act
= 70.5
°
C
⇒
Deviation, calculated according to
∆
T = T
set
– T
act
∆
T =
0.2
°
C
⇒
Entry of the corrected value
∆
T as
correction factor “
”
= +0.2
°
C
The temperature control is thus internally altered so that the
desired 70.7
°
C is also attained in the external system. The
temperature displayed at the circulator and that of the exter-
nal system now correspond with each other.