Supplement to the chapter ”Operation”
The cooling unit is switched-on separately with its own mains
switch 21 . Switch on the cooling unit only when cooling is
actually required. The compressor starts with a slight jerk.
If the DC5-K75 was already in operation before,
please switch on the unit only after the refrigera-
tion section has been switched off for at least 60
minutes. If switch on is attempted earlier, the com-
pressor will try to start several times but without
success. This could lead to burn-out of the com-
pressor motor windings.
The second compressor of the K75 is always switched on
automatically after a delay of 30 seconds.
This delay is for protection of the compressor.
When the unit is started under unfavourable envi-
ronmental conditions, the pressostat safety device
may respond after about 1 minute. If this happens,
please press the reset button 2 . Usually the unit
will then start correctly without going into fault sta-
tus again. However, if fault status appears several
times in succession, the unit is defective and must
be sent in for servicing.