23
29
!
Unpacking / Setting Up
10
6.3
Ambient conditions according to EN 61010
•
indoors, max. 2000 meters above sea level,
•
ambient temperature 5 ... 40
°
C,
•
relative humidity max. 80%/31
°
C (
→
50%/40
°
C)
•
excess voltage category II, contamination level 2
6.4
Resting time after transportation
As we can unfortunately not guarantee that our refrigerated
circulators are always transported according to our recom-
mendations (i.e. upright), lubrication oil can leak from the
compressor into the cooling circuit.
If the refrigerated circulator is started up whilst still in this
state, the compressor may be damaged to the lack of oil.
Therefore:
Rest the unit for 24 hours after setting up.
6.5
Ventilation
Keep all ventilation grids 23 and 29 free from obstruc-
tion to ensure unhindered air circulation.
Blocked ventilation grids lead to increased unit
heating which in turn reduces the cooling capacity
and thus impairs correct functioning.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com