34
.
.
34a
31
37
37
38
9
40
9a
40
!
Assembly
13
8.
Assembly
(if necessary after servicing)
1
Stand the temperature control module DC5 upside-
down,
Dry unit first if necessary, in order to avoid expos-
ing the electronics to water penetration.
2
Slide the seal 34 over the shaft up to shortly before the
screw heads 34a. Please note the the screws are not
symmetrically aligned!
Mounting the plate:
3
Slide the plate 37 with the accessories already fitted
over the shaft as far as the seal 34. Now slide it and the
seal as far as the screw heads or stopper.
4
Insert the screws 31 through the plate and screw tight.
Mounting the swivel part of the circulation set 40:
5
Preparation:
Unscrew the screw 9 from the unit shaft and loosen
screw 9a of the circulation set 40.
6
Locate swivel part 40,
7
Insert screw 9. Slightly tighten both screws 9 and 9a.
Mounting the plate onto the cooling unit K75:
8
Locate the plate 37 with the attached temperature con-
trol module on top of the bath vessel / cooling unit and
secure using the four sunken screws 38.
9
Fit the supplied conical bungs in the thermometer holes
in the bridge.
If disassembly is necessary, follow these steps in re-
verse order.