RS232C Interface for DC5
38
17.7 DC5 keys
The Enter key on the DC5 can be blocked if the circulator is
controlled via a PC. This prevents values set via the interface
from being accidentally altered at the circulator. The key
functions UP, DOWN and MENU remain available. Alter-
ations can however no longer be confirmed with the Enter
key.
L<cr>
Switching to REMOTE operation as well as blocking the En-
ter key (Lock)
U<cr>
Switching from REMOTE operation as well as releasing the
Enter key (Unlock)
17.8 Operating status
B<cr>
call up operating status:
” x
x
x
x
x
x
x
x ”, $<cr>,<lf>
0 : sensor OK
1 : sensor defective
0 : no alarm EXTERNAL
1 : alarm EXTERNAL
0 : no alarm pump
1 : alarm pump
0 : no alarm low liquid level
1 : alarm low liquid level
0 : no alarm excess temperature
1 : alarm excess temperature
0 : main relay present
1 : main relay missing
always 0, not used
0 : no active control
1 : active control