23
!
Draining
20
11. Draining
The temp. control unit is drained at the nozzle 23.
1
Place a suitable vessel underneath nozzle.
Bear in mind that the liquid will run out in a slight arc.
2
Turn plug slowly until it becomes disengaged from the
thread. A pin will prevent the liquid from running out
right away.
3
Pull out plug (pin) in one quick motion. The liquid will
start to run out.
4
Possible residues can be drained by tilting the circula-
tor slightly.
Hot heat transfer liquid should not be drained!
When certain conditions make draining necessary,
please act safety conscious: Wear protective
clothing and protective gloves!