RS232C Interface for DC5
36
17.3 Requirements made of external units
Only units which have been tested according to
EN 60 950 (=IEC 950) should be connected to the
DC5 interface!
17.4 Setting the desired set value
If the desired value is set via the computer, two horizon-
tal lines are shown on the display 5.
Additionally:
a) If the RS232C connection is interrupted, the circula-
tor controls the temperature to the last previously
set temperature value.
b) After the mains supply current has been switched
off and on, the circulator controls the temperature to
the last value entered via the DC5 keyboard itself
and not the value given by the computer
17.5 Correction value
If the correction value is set via the computer, two hori-
zontal lines are shown on the display 5. The correction
value only remains saved for as long as the supply volt-
age is switched on
Additionally:
a) If the RS232C connection is interrupted, the circula-
tor controls the temperature to the last previously
set temperature value.
b) After the mains supply current has been switched
off and on, the circulator controls the temperature to
the last value entered via the DC5 keyboard itself
and not the value given by the computer