8.3
Regular inspections / examinations
To keep the condition of the device in due order, examinations have to be performed regularly
according to the local laws and requirements of your country (e.g. in Germany based on BGV
A3, regulations for prevention of accidents as well as safety requirement inspections in
accordance with MDD Medical Device Directive, etc.).
For the next inspection date refer to the date on the inspection sticker on your device.
For h/p/cosmos sports and medical running machines and ladder ergometers, a maintenance interval respectively
technical safety checks of one year have been set. These examinations are only to be performed by trained and
authorized electricians.
The main inspection sticker on the device (e.g. running machine) also certifies the inspection of the optional equipment
and the accessories. However, inspection intervals for optional equipment and accessories (e.g. rope of h/p/cosmos
airwalk unweighting system, chest belt for safety harness, compressors, etc.) may deviate significantly from inspection
intervals of the main device. Read the manuals for details.
For all required measurements and control steps consider also local requirements of your country and see
detailed instruction and protocol form order no. [cos11690xx].
8.3.1
Visual inspections
N
Carry out visual inspection for damages of device and complete accessories: internal space, engine
compartment, connecting lead and the right position of the tensile relief and plug, ground wire connections,
etc.
N
Carry out visual inspection of mechanics and wear and tear parts: driving belt, tension roller, running belt,
lifting element with fixing bolts, welding seams at the frame, tight fitting of all screws and nuts. Follow the
appropriate maintenance list.
N
Clean device and engine compartment/internal space. Remove dust and dirt from the cooler openings of the
fan housing and the cooling air ducts of the drive motor, as well as from the ventilation slots and the cover of
the perforated plate of the frequency inverter.
N
Attach legible and complete hazard notes according to the instructions. Check the warning signs, protective
covers, step platform, engine hood and plastic cover installation channel in the engine compartment for
damage and replace if necessary.
N
All protective resistance connectors which can be reached from the outside must be examined for their correct
value and correct labelling.
N
Check capture slot at the rear end and at reverse belt rotation in the front as well and adjust if necessary.
Capture slot has to be < 8 mm, according to regulation draft 60601-2-xx © IEC:200X 62D/479/NWIP 2003-
05-18 and according to EN 957-1. See chapter entitled “Test finger“, which has a diameter of 9.5 mm (+ 0.1)