OP05
Maximum speed
The standard maximum speed of the selected device type (in
OPTION 03) will be indicated in the SPEED display.
Example: h/p/cosmos mercury 4.0 = 22.0 km/h; h-p-cosmos discovery
4.0 = 40 m/min
For safety purposes (e.g. for beginners), the max. speed of every
treadmill model can be limited.
=
indicates: maximums speed,
adjustable range from 1 km/h to maximum treadmill speed
selected unit is blinking,
max. is blinking
OP06
Maximum acceleration time in
seconds
(This option is only available for
treadmills, not for ladder
ergometers).
§
reports e.g. 131 (max. acceleration increment)
$
reports e.g. 5.0 (max. acceleration time in seconds.)
%
indicates: SEC.
OP07
Maximum elevation
(This option is only available for
treadmills, not for a ladder-
ergometer).
The standard maximum elevation of the selected device type (in
OPTION 03) will be indicated.
Example: h/p/cosmos mercury = 25 %
For safety purposes (e.g. for beginners, or in case the ceiling height
of the room's ceiling is insufficient) the max. elevation of every
treadmill model can be limited.
$
indicates: maximum elevation, adjustable range from 0 to
maximum treadmill elevation
OP08
Heart rate display interval
§
indicates heart rate display interval in seconds,
adjustable range from 0 (beat to beat) .... 1 to 9 seconds
Compare: An original POLAR heart rate monitor has a display interval
of 5 seconds
$
indicates: SEC
OP09
Test of all displays and LEDs
OP10
Enter the serial number of the
device provided on the nameplate
Change the values with
a
or
d
, c
onfirm the values with
c
Example: Treadmill with serial number cos30000va05-0128 on the
nameplate
§
indicates 300 (item number of device)
$
indicates: 0005 (item number of device)
%
indicates: 0128 (serial numbers of this device family)
OP11
Activating and de-activating speed
measurement
Speed measurement is normally activated / de-activated by selecting
the correct device type (administrator option number 3). Only in very
special cases a special adjustment in option number 11 is necessary.
Speed measurement can only be activated if the device has a speed
sensor.
§
indicates
0 = without speed measurement / -sensor
1 = with speed measurement / -sensor
(permanent control)
2 = with speed measurement / -sensor
(control as of speed > 2 km/h, no control < 2 km/h)