technical data
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 166 of 216
11.4
Running machines, 170-190/65 3p (quasar 3p / pulsar 3p)
pulsar
®
3p
locomotion
®
190/65 3p med
qu
as
ar
3p
qu
as
ar
3
p
lt
pu
ls
ar
3
p
pu
ls
ar
3
p
lt
lo
co
m
ot
io
n
19
0/
65
-3
p
E
m
ed
lo
co
m
ot
io
n
19
0/
65
-3
p
D
E
m
ed
UserTerminal
(display & keyboard)
6 LCD displays / 6 keys /
display of mode and unit with LED.
“lt-models“ and “3p E med” have no UserTerminal
(no display, no keyboard). Control via interface.
E: no Terminal
DE: UserTerminal TouchScreen E:
no Terminal
DE: UserTerminal TouchScreen
Power supply AC
(standard, read nameplate),
dedicated line required.
200 V 3N~ / 208 V 3N~ / 400 V 3N~
with PE
f: 50/60 Hz
Fuse (standard, read nameplate)
3 x C 16 A at 400 V 3N~
Power input (long time)
2700 VA
Power input (momentary)
5700 VA
(peak performances for milliseconds; max. 1s/min during heel strike)
Drive motor capacity
4300 VA
Elevation motor capacity
500 VA
Safety standards
n
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), VDE 0751,
IEC EN 60601-1-2, ISO 20957-1, EN 957-6,
Machinery Directive 2006/42/EC,
Earth leakage current
0.25 mA
Isolation transformer
3 x 1500 VA
Degree of protection
class I
M
/ type B
R
/ IP 20
Overvoltage category
Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt
Mode of operation
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
sports and medical: S,I,A (EN957)
Classification according to MDD
active therapeutic device and active diagnostic device, risk class II b
Speed
Definition
0.0 ... 40.0 km/h
0.1 km/h
0.0 ... 10.0 km/h
0.1 km/h
Speed accuracy*
+/- 5 % (class A based on EN 957)
Option special speed
0.0 ... 10.0 km/h or 0.0... 45.0 km/h
0.0 ... 25.0 km/h
Angle of inclination**
Definition
- 25.0 % ... + 25.0 %
0.1 %